Fuji

Thermal Imaging Systems

  Overview
The TP8 thermal imaging camera is used for plant condition monitoring, structural integrity inspection, quality control, and research and development in a wide range of industries - in a measurement range from -20 to 800 °C / -4 to 1470 °F.
On-board IR and visual cameras enable high resolution thermal and visual images to be stored simultaneously. Your preferred operational settings can be saved and recalled instantly. Automatic capabilities include hot spot and image centre detection. Large capacity (2 Gb) plus image sequence storage come as standard on the TP8.
     
Specification  
  • Imaging Performance
    Thermal
    Detector type: Uncooled FPA microbolometer (384 × 288 pixels)
    Spectral Range: 8-14 μm
    Thermal Sensitivity: 0.08 °С at 30 °С (Frame averaging algorithm)
    Frame Rate: PAL 50 Hz, NTSC 60 Hz
    Standard Lens Field of View/ Focus: 22°× 16°/ 35 mm
    Optional Lenses Field of View/ Focus: 7.7° × 5.8°/ 100mm; 45.6° × 35°/ 16mm
    Focus: Automatic or manual (motorized)
    Electronic Zoom: x1 to ×10 continuous zoom
    Visual
    Built-in Digital Video: CMOS Sensor, 1280 × 1024 pixels, 215 colors
  • Image Presentation
    External Display: 3.5″ high resolution color VGA LCD, 640 × 480 pixels
    Viewfinder: 0.6″ built-in high resolution color OLED, 640 × 480 pixels
    Video Output: VGA/PAL/ NTSC switchable
    Image Display: Thermal image alone/ Visual image alone/ Picture in picture
  • Measurement
    Temperature Range: -20 +800 °С / -4 to 1470 °F (up to 2000 °С / 3600 °F option)
    Accuracy: ±1 °С or ±1 % of reading
    Measurement Modes: Auto hot spot & auto alarm in live/zoomed image & video; 8
    moveable spots, 8 moveable & changeable areas displaying
    either max., min., or average, vertical & horizontal line profile,
    histogram & isotherm in live/zoomed/frozen/
    saved image & video
    Emissivity Correction: Variable from 0.01 to 1.00 (in 0.01 increment)
    Measurement Features: Automatic correction based on distance, relative humidity,
    atmospheric transmission and external optics
    Optics Transmission Correction: Automatic based on signals from sensors
  • Image Storage
    Type: Removable 2GB SD card or built-in flash memory
    File Format: JPEG (an individual file consists of infrared image,
    visual image, voice annotation and text annotation if any)
    Voice Annotation: Up to 30 seconds per file (more than 30 seconds optional)
    Bluetooth wireless headset
    Text Annotation: Selected from preset texts
  • Live Video Recording, Measurement and Storage
    Recording/Measurement: Thermal video recording to PC via USB 2.0
    Storage: In PC, capacity dependent on PC hard disk capacity
  • Laser Locator
    Classification Type: Class 2 semiconductor laser
  • Power Source
    Battery Type: Rechargeable Li-ion Camcorder battery, field-replaceable
    Charging System: In camera or in battery charger
    Battery Operating Time: Over 2.5 hours continuous operation
    External Power Operation: AC adapter 110 / 220 VAC, 50/ 60 Hz
  • Environmental Specification
    Operating Temperature: -20 to + 60 °С / - 4 to 140 °F
    Storage Temperature: -20 to + 60 °С / -4 to 140 °F
    Humidity: Operating and storing 10% to 95%, non-condensing
    Enclosure: IP54 IEC 529 housing
    Shock: Operational: 30G, IEC 68-2-29
    Vibration: Operational: 3G, IEC 68-2-6
  • Interfaces
    USB 2.0: Real-time data transfer and control via PC
    RS232 communication: Control of camera on PC
  • Operator Communications
    Touch Screen: Camera can be setup and operated via the touch screen
    Auto Speech Recognition: Automatically recognize and react to operator's voice
    commands
  • Physical Characteristics
    Housing: Magnalium
    Weight: 0.85 kg / 1.8 lb (excluding battery & LCD);
    1.1 kg / 2.4 lb (including battery & LCD)
    Size: 186 × 106 × 83 mm / 7.3 x 4.2 x 3.3 in (standard model)
    Tripod Mounting: 1/4 ″- 20