This is a paperless recorder that displays measured data on
the LCD in real time and stores data in CompactFlash.
The type of input such as thermocouple, resistance bulb,
D.C. voltage (current), etc. can be arbitrarily set to 36 channels
at the maximum.
The data stored in CompactFlash can be regenerated on the
screen, and the use of supplied support software allows the
data to be regenerated on a PC screen.
The data recorded in ASCII format can be directly read in a
spreadsheet such as Excel, which facilitates the processing
on a PC. (The data recorded in binary format cannot
be read in.) |
Number of input points:
9, 18, 27 or 36 points (Can be selected
at the time of purchase)
Input circuit: Input mutual isolation (See ”Others” on
page 4 for the withstand voltage)
Resistance bulb measured current:
about. 1 mA
Measuring cycles:
9 or 18 points….100ms cycles
27 or 36 points….200ms cycles
Recording cycle: 1 second to 12 hours
Input types: Thermocouple, resistance bulb, DC
voltage, and DC current (Shunt resistors
are fi tted in input terminals).
Note) Provide a shunt resistor (type:
PHZP0101) separately.
Selection of input types:
By key operation on the front panel. Note
that the same input type (thermocouple,
resistance bulb, voltage) should be set every
2 channels. Refer to "Setting method
of input types" for details.
Burn-out function:
Provided as standard for thermocouple
and resistance bulb inputs. If the input
has been open-circuited, the recording
level swings over 100%.
Thermocouple burn-out current:
approx. 0.2 μA
Input filter function:
Settable for each channel (primary delay
fi lter)
Time constants are settable in the range
from 0 to 900 sec.
Scaling function: Possible by DC voltage (current) input
Scaling range: -32767 to 32767
Decimal position:
settable at any point
Unit symbol: Selectable out of 125 different
units or 12 user
units of up to 7 characters.
Subtraction function:
Subtraction between each channel is
allowed.
Totalizing function:
The measured value of each channel
can be totalized. Applicable to daily,
monthly, annual or external input totalizing.
F value calculation function:
F value (extinction value of bacteria by
sterilization by heating) can be calculated
from the measured temperature
by each channel.
Square rooter function:
Square rooter can be performed
against the input value per each channel.
Computation function:
The following calculation is available with
the computation function.
(1) Computation function:
Addition, subtraction, multiplication,
division, absolute value, exponential,
square-root extraction, LOG, LN, EXP,
humidity, maximum, minimum, average,
and integration.
(2) Computation input enable:
Analog input (Ch1 to 72), integration
input (Ch1 to 72), DI (DI1 to 16), communication
input (No.1 to 36), and
constant number (No.1 to 60).
|