Performance Specifications:
Repeability of analysis : ± 0.05% of heat value
Functional Specifications:
Principle of measurement : Gas Chromotography
Measured gas streams: Up to 4 streams.
Analyzed components: 11
Analysis time (standart): 300sec
Detector: Micro TCD (Thermal Conductivity Detector)
Chromatographic methods:ISO 6974, part 4
Heat value calculation methods: GPA 2172
Gas to be analyzed: Natural gas
Auto-calibratorion: External solenoid valve & contact required
Ce Mark: Electromagnetic compatibity (89/336/EEC, 92/31/EEC, 93/68/EEC)
Analyzer output protocol: Foundation Fieldbus, MODBUS RTU or ASCII via RS232,485,422
Process Gas Conditions
Temperature: 14ºF to 122ºF(-10ºC to 50ºC)
Flow rate: 50 ± 20ml/min
Pressure: 7 to 70 psig measured at flow meter inlet
Dust and mist : None
Moisture : 2000 ppm or less
Coexisting component limits: H2,HE,O2,H2S(dry)<0.1 mol% each.
Ambient huminidy range: 0-95%RH
|