FujiFuji
Gas Chromatograph
  Overview
Transmitter Solution for NATURAL GAS ANALYSIS.

Compact and Lightweight Design

Easy Installation

Field Mountable at the Pipeline.

No Analyzer Shelter Required.

Eliminates Repairs in the Field.

Analysis of 11 Components in 300seconds.

Calculation of 20 Paramaters.

     
Technical Specification  

Performance Specifications:
Repeability of analysis : ± 0.05% of heat value

Functional Specifications:
Principle of measurement : Gas Chromotography
Measured gas streams: Up to 4 streams.
Analyzed components: 11
Analysis time (standart): 300sec
Detector: Micro TCD (Thermal Conductivity Detector)
Chromatographic methods:ISO 6974, part 4
Heat value calculation methods: GPA 2172
Gas to be analyzed: Natural gas
Auto-calibratorion: External solenoid valve & contact required
Ce Mark: Electromagnetic compatibity (89/336/EEC, 92/31/EEC, 93/68/EEC)
Analyzer output protocol: Foundation Fieldbus, MODBUS RTU or ASCII via RS232,485,422

Process Gas Conditions
Temperature: 14ºF to 122ºF(-10ºC to 50ºC)
Flow rate: 50 ± 20ml/min
Pressure: 7 to 70 psig measured at flow meter inlet
Dust and mist : None
Moisture : 2000 ppm or less
Coexisting component limits: H2,HE,O2,H2S(dry)<0.1 mol% each.
Ambient huminidy range: 0-95%RH


Documents:
HGC303 Brochures