Fuji
OMA-300 Process Analyzer
  Overview
The OMA-300  is a diode-array fiber-optics process spectrophotometer, measuring the absorbance of a sample as a function of wavelength. It is constructed of a light source, a dispersion device to select a particular wavelength, a sample area, and a detector
     
Technical Specification    

Repeatability:

± 0.5% of scale

Photometric accuracy:

± 0.004 AU

Operating conditions:

Temperature:

0 to 550C (320 – 1310F) standard -200 to 550C (-4 to 1310F) optional

 

Sample conditions

Immersion probe

Temperature

-200 to 1500C (-4 to 3020F)

Pressure

100 bar (1470 psig)

Flow through cell

Pressure

10 bar (147 psig) standard

Temperature

-200 to 1500C (-4 to 3020F)

Note

Wetted flow cell materials are: 316 stainless steel, K7 glass and Teflon. Other materials – optional

 

Area classification:

General purpose:

Standard

Class I Div. II Grp C&D

Optional

 Analog Output:

One galvanically isolated 4-20mA (additional 2 – optional)

Fault relay:

One SPDT alarm relay , to be set by user

Power:

80 to 240 Volts AC 40 to 60Hz, 40 Watts.

Physical Dimensions: 
 

Dimensions

16"(H) x 14" (W) x 8" (D) (41 x 36 x 20 cm)

Weight

32 lbs. (15 Kgs.)

   
Documents PB OMA-300 R0810.pdf